Calculates commonly used indices from absorbance and excitation–emission matrix (EEM) data. Also checks and flags index values for potential issues.
Usage
get_indices(
eemlist,
abslist,
index_method = "eemanalyzeR",
tolerance = 0.2,
return = "wide",
cuvle = 1,
qaqc_dir = NA,
arg_names = NULL
)Arguments
- eemlist
An
eemlistobject.- abslist
An
abslistobject.- index_method
Either "eemanalyzeR", "eemR", "usgs", or a custom function. See Details.
- tolerance
Maximum percent deviation that the check standard can vary from the long-term values without being flagged.
- return
Output format: "long" or "wide".
- cuvle
Cuvette (path) length in cm.
- qaqc_dir
File path to the QAQC files generated with
create_mdl()andcreate_std(). Default is a user-specific data directoryrappdirs::user_data_dir().- arg_names
Optional list of arguments passed from higher-level functions for README generation.
Value
A list with two data frames:
eem_index: fluorescence indices
abs_index: absorbance indices
Details
Index methods
Three preset index sets are available:
eemanalyzeR:
eemanalyzeR_indices()eemR:
eemR_indices()usgs:
usgs_indices()
You may also pass a custom index-generating function. See
vignette("custom-indices") for
instructions.
QA/QC flags
Index values are checked for common issues. Flags include:
DATA01: Missing data required for calculation
DATA02: Missing required wavelengths; value may be inaccurate
DATA03: Ratio denominator was zero
DATA04: Spectral slope could not be calculated
DOC01: Missing dissolved organic carbon (DOC) data
INF01: Infinite value
MDL01: All values below MDL
MDL02: One or more values below MDL; use cautiously
MDL03: Ratio index where numerator or denominator was entirely below MDL
NEG01: Negative value
STD01: Check standard value outside tolerance
VAL01: Value below typical range
VAL02: Value above typical range
Examples
indices <- get_indices(
example_processed_eems,
example_processed_abs,
qaqc_dir = system.file("extdata", package = "eemanalyzeR")
)